Refinement of Moisture Calibration Curves for Nuclear Gage: Interim Report no. 1

Report No: 71-R28

Published in 1972

About the report:

This study was initiated to determine the correct moisture calibration curves for different nuclear gages. It was found that the Troxler Model 227 had a linear response between count ratio and moisture content. Also, the two calibration curves for this gage were best divided by the optimum moisture content of the soil. The separation value of optimum moisture content was 21%. The Troxler Model 2401 was found to need only one calibration curve to produce a standard error of 1.7 pcf.

Disclaimer Statement:The contents of this report reflect the views of the author(s), who is responsible for the facts and the accuracy of the data presented herein. The contents do not necessarily reflect the official views or policies of the Virginia Department of Transportation, the Commonwealth Transportation Board, or the Federal Highway Administration. This report does not constitute a standard, specification, or regulation. Any inclusion of manufacturer names, trade names, or trademarks is for identification purposes only and is not to be considered an endorsement.